Electrochemical Corrosion Measurements

Electrochemical Corrosion Measurements Primer

Most metal corrosion occurs via electrochemical reactions at the interface between the metal and an electrolyte solution. A thin film of moisture on a metal surface forms the electrolyte for atmospheric corrosion. Wet concrete is the electrolyte for reinforcing rod corrosion in bridges. Although most corrosion takes place in water, corrosion in non-aqueous systems is not unknown.

Corrosion normally occurs at a rate determined by an equilibrium between opposing electrochemical reactions. The first is the anodic reaction, in which a metal is oxidized, releasing electrons into the metal. The other is the cathodic reaction, in which a solution species (often O2 or H+) is reduced, removing electrons from the metal. When these two reactions are in equilibrium, the flow of electrons from each reaction is balanced, and no net electron flow (electrical current) occurs. The two reactions can take place on one metal or on two dissimilar metals (or metal sites) that are electrically connected.

Basics of electrochemical corrosion measurement

Figure 1-1. Corrosion Process Showing Anodic and Cathodic Current Components.

Figure 1-1 diagrams this process. The vertical axis is potential and the horizontal axis is the logarithm of absolute current. The theoretical current for the anodic and cathodic reactions are shown as straight lines. The curved line is the total current — the sum of the anodic and cathodic currents. This is the current that you measure when you sweep the potential of the metal with your potentiostat. The sharp point in the curve is actually the point where the current changes signs as the reaction changes from anodic to cathodic, or vice versa. The sharp point is due to the use of a logarithmic axis. The use of a log axis is necessary because of the wide range of current values that must be displayed during a corrosion experiment. Because of the phenomenon of passivity, it is not uncommon for the current to change by six orders of magnitude during a corrosion experiment.

[…]

Electrochemical Quartz Crystal Microbalance - Gamry Instruments's eQcm

Quartz Crystal Microbalance

The Quartz Crystal Microbalance (QCM) is an exciting tool for the electrochemist. With it, the researcher can now follow not only the current that flows, but the weight changes of the electrode, too! This is a valuable tool when studying reactions which involve films, adsorbates, metal deposition, corrosion, or monolayer formation. It is sensitive enough Read more about Quartz Crystal Microbalance[…]

cyclic voltammetry scan rate

How Fast a CV Can I Measure Reliably ?

Being able to scan rapidly does not insure that the results will be meaningful! The speed of the current measurement circuitry is often the limiting factor!

The key to finding the practical limit for obtaining meaningful fast scan cyclic voltammograms is nearly always finding the speed of the current measurement. Here the researcher has an important role to play: It is the researcher who must select the current range to use for fast cyclic voltammetry. The autoranging capability of many modern computer controlled potentiostats generally cannot be used because the decisions cannot be made and implemented fast enough.

Because of stray (and deliberately added) capacitances, the current measuring circuitry generally becomes slower as the full scale current decreases. Obtaining the fastest scan requires a tradeoff of scan rate, electrode size, analyte concentration, current range, and acceptable noise in the measurement. It is often better to use a less sensitive current scale (larger full scale current) coupled with a larger pre-amplification factor on the ADC, data recorder, or oscilloscope used. Although this approach is likely to increase the noise in the measurement, it does allow a higher scan rate to be realized.

The speed or frequency response of each current range can sometimes be found in the manufacturer’s data sheet under “Current Measurement” or sometimes as a “System Specification” if a specific current range is quoted along with the bandwidth.

This number can be roughly translated into a scan rate by looking at Figure 1.

A linear scan for cadmium with a sine wave superimposed

[…]

There is a description of potentiostat stability (written by DK Roe) in the Kissinger & Heineman book (

How Fast Can My Potentiostat Scan?

Several times I have been asked a question about whether potentiostat Model XYZZY can scan at xyz V/s. Rarely is the answer in the data sheet for the instrument. Often, however, enough information is given to assess the limits. The path I follow to arrive at an answer is outlined here. An important fact about Read more about How Fast Can My Potentiostat Scan?[…]

There is a description of potentiostat stability (written by DK Roe) in the Kissinger & Heineman book (

Fitting EIS Data – Adding Components

Dr. Bob on  EISOne guideline that I have heard recommended (although I cannot give a reference for it) is that data over a decade range of frequency is required to support each circuit component.

All curve-fitting software should report some measure of the “goodness of fit.” Often this is the chi-squared parameter ( X2 ) or a value related to it. Boukamp makes the recommendation that the value of X2 should decrease by tenfold if a new circuit element is introduced into the circuit model. The tenfold decrease provides the justification for including the new circuit element. If the inclusion of an additional circuit element does not substantially improve the goodness-of-fit (as evidenced by the decrease in the X2 value), then based on Occam’s Razor, you should keep the simpler model, or continue your search for an improved one.

The old joke about the ability to “fit an elephant” if you use enough parameters is all too true with impedance data. Each component added to the model should have a physical explanation. Adding components only because they make the fit look better (smaller X2) without a physical interpretation is the equivalent to “fitting an elephant.”

What is X2?

[…]

Development of a Cyclic Voltammetry Method

For DNA Electrochemical Detection on Microfluidic Gene Chip On the microfluidic gene chip, due to high difficulty in temperature changes frequently and products detecting equipment miniaturize, the conventional methods of DNA detection can’t meet the requirements. In this paper, a newly electrochemical method, cyclic voltammetry, basing on a set of special electrodes and the Loop-mediated Read more about Development of a Cyclic Voltammetry Method[…]

National Science Foundation

NSF Creates Electrochemical Research Center

NSF Creates industry electrochemical research center at Ohio University

The Center for Electrochemical Engineering at Ohio Univ.’s Russ College of Engineering and Technology has received a National Science Foundation (NSF) award to establish a new industry university cooperative research center in Athens, Ohio, with partner site Washington University-St. Louis.

Led by Russ Professor of Chemical Engineering and Center for Electrochemical Engineering Director Gerri Botte, research at the new Center for Electrochemical Processes and Technology (CEProTECH) will focus on electrochemical alternatives to conventional chemical and biological processes, with the goal of enhancing advanced production capabilities, via a consortium model.

Consortium members will have access to pre-competitive, industry-driven research results and a dedicated 20,000-square-foot facility, located on Mill Street in Athens, Ohio, with more than $7 million in state-of-the-art equipment and infrastructure; students with specialized expertise in electrochemical engineering; and relationships with faculty, government labs and agencies, and other industry members.

[…]

Electrochemistry Impedance Spectroscopy Basics

Basics of Electrochemical Impedance Spectroscopy

This application note presents an introduction to Electrochemical Impedance Spectroscopy (EIS) theory and has been kept as free from mathematics and electrical theory as possible. If you still find the material presented here difficult to understand, don’t stop reading. You will get useful information from this application note, even if you don’t follow all of the discussions.

Four major topics are covered in this Application Note.

  • AC Circuit Theory and Representation of Complex Impedance Values
  • Physical Electrochemistry and Circuit Elements
  • Common Equivalent Circuit Models
  • Extracting Model Parameters from Impedance Data

No prior knowledge of electrical circuit theory or electrochemistry is assumed. Each topic starts out at a quite elementary level, then proceeds to cover more advanced material.

AC Circuit Theory and Representation of Complex Impedance Values

Impedance Definition: Concept of Complex Impedance

concept of electrical resistance

(1)

Almost everyone knows about the concept of electrical resistance. It is the ability of a circuit element to resist the flow of electrical current. Ohm’s law (Equation 1) defines resistance in terms of the ratio between voltage, E, and current, I.

While this is a well known relationship, its use is limited to only one circuit element — the ideal resistor. An ideal resistor has several simplifying properties:

  • It follows Ohm’s Law at all current and voltage levels.
  • Its resistance value is independent of frequency. AC current and voltage signals though a resistor are in phase with each other.

[…]

Reference Electrodes

Introduction

This Application Note presumes that you have a basic understanding of potentiostat operation. If you are not that knowledgeable concerning electrochemical instrumentation, please read Potentiostat Fundamentals before continuing. Experienced potentiostat users may skip the primer and read on.

It’s only natural that electrochemists concentrate on the working electrode. After all, reactions at the working electrode are Silver Chloride Reference Electrodewhat they study. However, the reference electrode shouldn’t be ignored. Its characteristics can greatly influence electrochemical measurements. In some cases, an apparently “good” reference electrode can cause a complete failure of the system.

For reliable reference electrode performance, you should assign a “Lab Master” and treat it very, very carefully so it can serve as a standard for your other reference electrodes. Never use the Lab Master in an actual experiment. The only purpose of the Lab Master is to serve as a check for the other reference electrodes. If a reference electrode is suspected to be bad, you can check the potential versus the Lab Master. You can do that with a voltmeter, or with your Gamry Potentiostat by running and open circuit potential. If the potential difference is less than 2-3 mV, it’s OK. If it’s higher than 5 mV, it needs to be refreshed or discarded.

[…]

Gamry Interface - Reference Potentiostats

Potentiostat Fundamentals

Introduction

A potentiostat is an electronic instrument that controls the voltage difference between a Working Electrode and a Reference Electrode. Both electrodes are contained in an electrochemical cell. The potentiostat implements this control by injecting current into the cell through an Auxiliary or Counter electrode.

In almost all applications, the potentiostat measures the current flow between the Working and Counter electrodes. The controlled variable in a potentiostat is the cell potential and the measured variable is the cell current.

Gamry Potentiostat

Prerequisites

This Application Note may be difficult to follow unless you have some familiarity with electrical terms such as voltage, current, resistance, frequency, and capacitance. If you feel your knowledge in this area is lacking, we suggest review of a very basic electronics or physics book.

Electrodes

A potentiostat requires an electrochemical cell with three electrodes as shown below. W/WS denote the working and working sense. R denotes the reference electrode and C denotes the counter electrode.

Potentiosta: Electrochemical cell-3 electrodes […]

measuring surface related currents

Measuring Surface Related Currents

Using Digital Staircase Voltammetry

Introduction

Cyclic Voltammetry (CV) is unarguably the most popular electrochemical technique. It owes its well deserved reputation to its ability to deduce reaction mechanisms with relatively low cost equipment and quick experimentation. Since the very highly cited paper by Nicholson and Shain1 the technique has been the centerpiece of any electrochemical study.

CV involves sweeping the potential linearly between two limits at a given sweep rate while measuring current. The sweep rate chosen can be varied from few microvolts per second to millions of volts per second.

Electrochemical instrumentation has evolved vastly since the days of Nicholson and Shain. Currently, most manufacturers (including Gamry) make digital instruments with digital signal generators. These signal generators approximate the linear sweep with a staircase of variable step sizes and durations.

CV Electrochemical Technique

Figure 1 Staircase vs analog ramp

[…]

Testing Electrochemical Capacitors

Cyclic Voltammetry and Leakage Current

Purpose of This Note

This application note is the first part of an overview of electrochemical techniques used to test electrochemical capacitors (ECs). Electrochemical capacitors that are commercially available were tested to explain and discuss the theoretical background of cyclic voltammetry and leakage current measurement.
testing capacitors: cyclic voltammetry

Introduction

This application note is part one of 3.  Part 2 of this note discusses techniques that are also familiar to battery technologists. Part 3 describes theory and practice of EIS measurements on capacitors.

In contrast to batteries, ECs generally store energy by highly reversible separation of electrical charge while batteries use chemical reactions. ECs consist of two high-surface electrodes immersed in a conductive liquid or polymer called the electrolyte. The electrodes are separated by an ionic-conducting separator that prevents shorts between the two electrodes.

[…]